Home > Products > Biosafety Laboratory > Airtight Door
Return to list

Airtight Door

Airtight Doors have very high airtightness and are mainly used in rooms within high-level biosafety laboratories that require airtightness to protect the airtightness of the biological laboratory enclosure.

  • Series: KuDo

  • Compliance: Third-party test report obtained
  • Structure: Fully welded construction for excellent airtightness



Introduction

Airtight Door

Airtight Doors have extremely high airtightness, effectively preventing gas infiltration. They are typically used to control gas flow and maintain pressure balance. They are mainly used in high-level biosafety laboratories, isolation rooms, sterilization rooms, and other rooms with high airtightness requirements to protect the airtightness of the biological laboratory enclosure.

Airtightness

There are two sealing methods for biological airtight doors: air-filled sealing and mechanical compression sealing.

The inflatable sealing biological airtight door achieves locking and sealing through an electromagnetic lock and an inflatable sealing strip: an inflatable sealing strip is installed in the groove structure on the side of the door panel. When the door is closed, the controller senses it, and the inflatable sealing strip begins to inflate and expand, achieving sealing and locking between the door panel and the door frame; when the door is opened, the inflatable sealing strip deflates and retracts into the groove, and the door panel and door frame are in the unlocked state.

Mechanically clamped biological airtight doors achieve locking and sealing through mechanical clamping and an electromagnetic lock: Strong silicone sealing strips are installed on the sides of the door panel to achieve clamping and sealing, ensuring airtightness. During normal use, pressing the door open button unlocks the electromagnetic lock, opening the mechanical clamping mechanism to open the door; the method of use is the same for both inside and outside. When two doors are in use, they can be interlocked.

Protective

The airtight door has excellent airtightness, effectively preventing the leakage and spread of biological contaminants through physical isolation, thus protecting personnel safety. The entire structure is made of fully welded stainless steel, making it corrosion-resistant, easy to clean, robust, stable, and durable.

Equipped with an electromagnetic lock, when both doors are used simultaneously, the electromagnetic lock can interlock the inner and outer doors; when a single door is used, the electromagnetic lock can lock twice to ensure safety.

The visual design features a double-layered tempered glass window that is strong, corrosion-resistant, and provides a better user experience.

Advantages of Biosafety Airtight Doors

All stainless steel fully welded structure

Rounded corner structure to avoid dead corners in sealing

Double tempered glass window

Electromagnetic lock, good expandability

High-density corrosion-resistant inflatable sealing strips/silicone sealing strips

Emergency stop button and manual vent valve ensure safety.

Third-party testing report

Inspection basis:

The testing methods and evaluation standards were conducted in accordance with GB50346-2011 "Technical Specification for Biosafety Laboratory Building", RB/T199-2015 "Technical

Specification for Biosafety Performance Evaluation of Laboratory Equipment" and the entrustment contract.

Air tightness test results:

The airtightness of a sealed space (1830mm*1180mm*2620mm) formed by the connection between the airtight door and the enclosure structure was tested using the pressure decay method. Verification showed that the airtightness of the sealed space, under an initial pressure of -500Pa, met the laboratory specification requirement that the pressure not fall below -250Pa after 20 minutes. The tested sealed space also met the commissioned requirement that, at a pressure 1000Pa lower than the surrounding environment, the amount of air leaking per minute should not exceed 0.1% of the chamber volume.

Performance test results of inflatable and airtight doors:

The switch buttons and status indicator lights function normally. When opening the door, the inflatable sealing strip automatically deflates; after deflation, the electromagnetic lock disengages, and the indicator light illuminates red. When closing the door, the electromagnetic lock closes, the inflatable sealing strip automatically inflates, and the indicator light illuminates green after inflation is complete. When the door is closed, pressing the emergency unlock switch allows the door to open; opening the emergency deflation valve deflates the inflatable sealing strip, allowing the door to open. The inflation and deflation times of the sealing strip are both less than 5 seconds.

Performance test results of mechanically clamped airtight doors:

When performing door opening, closing, and locking operations, the moving mechanism functions normally, and the electromagnetic lock and door lock switch function normally. The switch buttons and status indicator lights function normally; during a normal door opening operation, the electromagnetic lock disengages and the indicator light illuminates red; during a door closing operation, the electromagnetic lock closes and the indicator light illuminates green. When the door is closed, pressing the emergency unlock switch will allow the door to open.

Currently, most existing P4 laboratories in China use inflatable airtight doors, although some cases use mechanically compressed airtight doors. All of these meet the requirements for airtightness testing using the pressure decay method for the enclosure structure. Existing large animal P3 laboratories generally use mechanically compressed airtight doors in their protected areas, which meet the requirements for airtightness testing using the constant pressure method for the enclosure structure. (Refer to "Design and Construction of Biosafety Laboratories")

Technical Specifications

Inflatable type airtight doors achieve doorTechnical Parameters

ItemParameters
Main materialSUS 304
Structural pressure resistance≤1500 Pa
Inflation time<5sec
Deflation time<5sec
Inflation/deflation cycles10000 cycles (no rupture or leakage)
Compressed air3-4 bar
Power supplyAC 220V/50Hz
Operating environmentTemperature:-30C~50°C; Relative humidity: <90%
Air-tightnessNatural pressure decay ≤50% after 20 min at -500

Note: Dimensions and other functions can be customized.


Mechanical compression type airtight doorTechnical Parameters

ItemParameters
Main materialSUS304
Structural pressure resistance≤1500Pa
Door operation methodManual door operation, 3-point compression locking
Interlock methodElectromagnetic interlock
Sealing stripSilicone foam gasket
ControlEmergency stop, indicator lights
Power supplyAC220V/50Hz
Operating environmentTemperature: -30°C~50°C; Relative humidity: <90%
Air-tightnessNatural pressure decay ≤50% after 20min at -500Pa

Note: Dimensions and other functions can be customized