Portable Exhaust Air Cleaner
The Disinfection Exhaust Air Cleaner is a self-circulating high-efcien-cy air filtration device. It can function as a positive-pressure clean air circulation system to continuously purify indoor air. Alternatively, it can operate as a local or full exhaust system to create a negative-pressure isolation room, rapidly converting standard hospital wards into compli-ant negative-pressure isolation wards.
•Series:KuHo
Introduction

Exploded View Diagram
1.Air Outlet 2 .Hepa Filter Locking Mechanism
3.Hepa Filter Differential Pressure Gauge 4.Power Switch
5.Speed Controller Interface 6.Power Supply Port
7.Uv Lamp 8.Pre-filter Differential Pressure Gauge
9.Air Intake Grille 10.Casters (Or Universal Wheels)
11.Blower (Or Fan) 12.Blower Guard (Or Fan Cover)
13.Hepa Filter 14.pre-filter
Product Advantages
Self-circulating design enables rapid provision of localized clean microenvironments.
Visualized control with touch screen and PLC for adjustable air purifier airflow.
Connectable to speed controllers for adjustable wind speed and room air exchange rate modification.
Configurable automatic power on/off timing.
Three-sided air intake and outlet design minimizes purification dead Zones.
Dual differential pressure gauges monitor pre-filter and HEPA filter pressure drops, prompting timely replacements.
Portable universal wheels allow free indoor mobility.
Equipped with industrial-grade maintenance-free EBM blowers.
Working Principle
Place the disinfection exhaust air purifier in the room requiring purification and power on the unit. Air enters through the bottom intake grille, passes through the pre-filter and HEPA filter, then is discharged as clean air at high velocity from the top outlet grille. This continuous circulation filtration achieves rapid indoor air purification.

Technical Specifications
| Maximum Airflow | 1500 CMH | 2500 CMH |
|---|---|---|
| Dimensions (W * D * H) | 680*315*1900 mm | 745*700*1700 mm |
| Rated Power | 265W | 350W |
| Weight | 65 kg (excluding filter) | 105 kg (excluding filter) |
| Air Outlet Diameter | 200 mm | 400 mm |
| UV Lamp Power | 14 W | 14 W |
| UV Lamp Lifespan | 8000 hours | 8000hours |
| HEPA Filter Efficiency | H13 (EN1822) | H13 (EN1822) |
| Filter – Recommended Replacement Cycle | Pre-filter: 1 month HEPA filter: 12 months | Pre-filter: 1 month HEPA filter: 12 months |