Home > Products > Pharmaceutical Industry > Negative Pressure Isolator
Return to list

Negative Pressure Isolator

The KuNe series negative protective isolators maintain a negative pressure state within a sealed cavity, controlling the unidirectional airflow from the outside to the inside. A high-efficiency filtration system processes the exhausted air, preventing the leakage of cytotoxic and highly reactive chemicals, thus ensuring operator and environmental safety.
Applications: Sampling, weighing and dispensing, powder mixing, tableting, coating, feeding, and weighing and packaging of finished API materials in the pharmaceutical industry.

  • Series: KuNe

  • Optional: glove leak tester, α-B valve, RTP rapid transfer port

  • Filter: cartridge-type PUSH-PUSH HEPA filter with contactless replacement

  • Airflow direction: vertical unidirectional airflow

Introduction

※ The enclosure is made entirely of stainless steel, with a tempered glass door and an inflatable sealing strip to ensure the airtightness of the operating environment and prevent leakage of highly active, highly toxic, and highly allergenic materials; the interior of the enclosure is equipped with a dustproof and waterproof socket and a high-pressure water gun; optional accessories include a glove leak detector, an α-B valve, and an RTP quick transfer interface.

※ A double-door air-sealed interlocking device is used between the chambers to protect the environmental conditions of high-risk operating areas. A cartridge-type push-push high-efficiency filter is used, allowing for contactless filter replacement and protecting operator safety.

※ Vertical unidirectional airflow, with air flowing from top to bottom, ensures uniform airflow. The pressure within the chamber can be manually adjusted according to operational needs to prevent leakage of toxic substances and protect operator safety.

※ Emergency mode: When the isolator's seal or glove breaks, the pressure inside the chamber will rise rapidly. The isolator will automatically activate the alarm, adjust the fan frequency, increase the exhaust volume, and close the air inlet valve to ensure that the air velocity at the gaps is above 0.5 m/s, so as to ensure that highly active substances will not leak out and protect the safety of the operator and the laboratory environment.

Technical Specifications

ItemParameters
Filter EfficiencyH14 (EN1822)
Enclosure MaterialSUS 304
Protection ClassOEB4 or OEB5
Noise Level<65 dB (A)
Air Changes Per Hour (ACH)20-100 ACH
Differential Pressure GaugeDWYER (Digital Display)
PLC (Programmable Logic Controller)Siemens
Touch ScreenSiemens
Power SupplyAC 220 V/50Hz

Note: Dimensions and other functions can be customized.