Home > Products > Pharmaceutical Industry > Glove leak tester
Return to list

Glove leak tester

The portable sterile glove leak tester utilizes the pressure decay method (pressure drop method)and is a high-end portable model that complies with FDA,GMP,and ISO 14644-7 standards.

Introduction

Fully Automatic Pressure Decay Test Method

1.Rapid inflation – The instrument, equipped with an oil-free silent air pump,quickly inflates the sealed glove until the preset test pressure (300–3000 Pa)is reached

2.Stabilization (dwell phase)–Inflation stops,and the system enters a brief stabilization period to eliminate pressure errors caused by airflow fluctuations and glove deformation .

3.Leak test – A high-precision sensor continuously monitors pressure changes.If the glove has a pinhole or break,gas leakage will cause the pressure to drop.

4.Result determination – The instrument automatically calculates the pressure drop per unit time,compares it with a preset threshold,and directly displays PASS or FAIL.

Technical Specifications

ItemParameters
Power (W)240
Inflation pressure (Pa)300~3,000
Instrument weight (kg)7.8
Built-in pressure (MPa)0.2~0.6
Pressure display resolution (Pa)0.1
Power supplyBuilt-in battery
Battery cycle life30 times, 5-minute test
Power sourceAC 220V, 120W
Test time (min)3~10
DA(MP) (μm)400
Dimensions (mm)Disc diameter 310, 8
Online dataCan be stored
Printing functionUSB, Bluetooth (optional)
Data transmission methodUSB, Bluetooth
Test port specification (mm)Φ230 (customizable)
Host configurationComes with its own air source and battery