VHP Sterilizer
The VHP sterilizer injects H₂O₂ solution into a closed-loop chamber. Utilizing advanced low-temperature flash vaporization technology, the H₂O₂ solution rapidly transforms into gas, which is quickly and evenly distributed throughout the sterilization space. The H₂O₂ gas rapidly kills spore-forming bacteria within the sterilization space, eliminating viable microorganisms and meeting GMP requirements. This sterilizer is widely used in laboratories, pharmaceutical engineering, and medical fields, and can be used in conjunction with isolators, pass-through windows, bag-in-bag-out systems, biosafety cabinets, and other equipment, providing a safe and reliable sterilization solution for high-level biosafety environments.
Series:KuVi
Introduction

Sterilization Process
DehumidificationThe HEPA filter controls the humiditywithin within a predetermined range, preparing the H₂O₂ gas concentration to remain below saturation (exposure)levels . adjust
H₂O₂ gas circulates rapidly in a closed environment.The settling time is affected by the sterilizing agent injection rate, sterilizer volume,material , and temperature.
Sterilization
Sterilization begins when the preset optimalsterilization , and continues until the preset sterilization effect value is achieved.
Emission
After sterilization is completed, the injection of H₂O₂ solution is stopped. The HEPA filter catalyzes the VHP generator to degrade the H₂O₂ gas in the chamber to meet emission standards.

Technical Specifications
| Item | Parameters |
|---|---|
| Material | SUS 304/SUS 316 |
| Power Supply | AC220V/50Hz |
| Power | 2.5Kw |
| Sterilization Cycle Time | ≥30min |
| Sterilization Process Temperature | ≤50℃ |
| Sterilization Process Humidity | 40%-70% |
| Sterilizing Agent | 35%H2O2 |
| Hydrogen Peroxide Vaporization Rate | 5-30ml/min |
| Sterilization Efficiency | log6 |
| Noise Level | ≤75dB (A) |
| Sterilizing Agent Residue | <1ppm |